Palomar Cynosure Repair Service – Systems we can Repair, Parts and Error Codes:
SYSTEMS
- Icon, StarLux 500 and StarLux 300
- Vectus
- Q5
ERROR CODES
Some Common error repair codes for Palomar/ Cynosure:
006- Remote interlock on back of the system is not plugged in
007- Emergency stop button is not plugged in
009- Water leak in hand piece
010- wiring in hand piece is loose
022- handpiece cooltip has water restriction
023- check water level in water tank
024 – Flashlamp water flow is restricted/low
044- internal PFC out put voltage exceeds 400 VDC +/- 10V
046- temperature exceeds 120 degrees
053 – Check handpiece for proper connection- all 4 connections are engaged
054- Flashlamp water pressure is too high
068- system is not able to establish simmer
069- internal lamp controller voltage exceeds limit
073 – is a broken wire which can be repaired
PARTS WE CAN PROVIDE:
- Water Pump
- Pressure Switch
- Flow Switch
- Lamp
- IPL Shell
- Screen
- Connectors